한국해양대학교

Detailed Information

Metadata Downloads

폴라로그래픽 전극법을 이용한 고정도 용존산소 측정시스템 개발에 관한 연구

Title
폴라로그래픽 전극법을 이용한 고정도 용존산소 측정시스템 개발에 관한 연구
Alternative Title
(A) Study on the Development of a High Accuracy Dissolved Oxygen Measuring System using Polarographic Method
Author(s)
尹甲烈
Issued Date
2001
Publisher
한국해양대학교 대학원
URI
http://kmou.dcollection.net/jsp/common/DcLoOrgPer.jsp?sItemId=000002174052
http://repository.kmou.ac.kr/handle/2014.oak/10467
Abstract
Dissolved oxygen in pure water supplied to power plant may cause corrosion of piping arrangements, condensers or turbine's blades. For preventing accidents from corrosion, therefore, it is essential to measure the concentration of Dissolved oxygen in real-time.


In this paper we present a method of measuring dissolved oxygen very accurately up to ppb units. This method, called polarographic method, is based on the measures of the electric current generated by the oxidation process in cathode and deoxidation process in anode, assuming that the amount of the current is proportional to the density of dissolved oxygen. We introduce algorithms for the compensation of temperature and atmospheric pressure to reduce measuring errors and multiple amplification ratio algorithm to amplify small current.


Effectiveness of the suggested method is verified through a series of experiments in the real power plant field.
Appears in Collections: